SpikeGLX is acquisition software for Neuropixels probes. This page shows how to convert one SpikeGLX session to Neurodata Without Borders (NWB) with NeuroConv, the open-source conversion library that CatalystNeuro maintains. The code and behavior described are as of NeuroConv 0.10.2, and the scripts were run against SpikeGLX test recordings.
What SpikeGLX Writes to Disk
SpikeGLX saves a run in a folder named after the run and its gate
index, such as my_run_g0. Every data file comes as a pair: a .bin
file holding the samples as 16-bit integers with no header, and a
.meta text file that records the sampling rate, channel map, gains,
and probe. File names carry the run name, the gate index (g0), the
trigger index (t0), and the stream.
my_run_g0/
├── my_run_g0_t0.nidq.bin
├── my_run_g0_t0.nidq.meta
└── my_run_g0_imec0/
├── my_run_g0_t0.imec0.ap.bin
├── my_run_g0_t0.imec0.ap.meta
├── my_run_g0_t0.imec0.lf.bin
└── my_run_g0_t0.imec0.lf.meta
The per-probe subfolders appear when the “Folder per probe” option is on. With it off, all files sit directly in the run folder. NeuroConv reads either layout. The streams are:
imecN.ap: the action potential band of probe N, nominally 30 kHz.imecN.lf: the local field potential band of probe N, at 2.5 kHz. Neuropixels 1.0 probes save the two bands as separate files. Neuropixels 2.0 probes are full-band and save a single.apfile.nidq: the analog and digital channels of a National Instruments device, typically task TTLs, camera frame pulses, and other rig signals.
Each probe file also carries a sync channel (SY), a status word in
which one bit is the square wave SpikeGLX uses to synchronize streams.
Installing NeuroConv
NeuroConv keeps the dependencies of each format in an optional extra:
pip install "neuroconv[spikeglx]"
To add Kilosort or Phy output as well, install
"neuroconv[spikeglx,kilosort]" or "neuroconv[spikeglx,phy]".
Converting One Session
SpikeGLXConverterPipe takes the run folder and writes every stream in
it to one NWB file.
from zoneinfo import ZoneInfo
from neuroconv.converters import SpikeGLXConverterPipe
folder_path = "path/to/my_run_g0"
# For example: ['imec0.ap', 'nidq', 'imec0.lf', 'imec0.ap-SYNC', 'imec0.lf-SYNC']
print(SpikeGLXConverterPipe.get_streams(folder_path=folder_path))
converter = SpikeGLXConverterPipe(folder_path=folder_path)
# Everything NeuroConv could read from the .meta files
metadata = converter.get_metadata()
# SpikeGLX records the start time without a time zone, so attach the
# time zone of the rig.
session_start_time = metadata["NWBFile"]["session_start_time"]
metadata["NWBFile"]["session_start_time"] = session_start_time.replace(
tzinfo=ZoneInfo("America/New_York")
)
metadata["NWBFile"]["session_description"] = (
"Head-fixed recording during a visual discrimination task."
)
metadata["NWBFile"]["experimenter"] = ["Last, First"]
metadata["NWBFile"]["lab"] = "My Lab"
metadata["NWBFile"]["institution"] = "My University"
# Subject metadata is required for upload to DANDI
metadata["Subject"] = dict(
subject_id="M001",
species="Mus musculus",
sex="M",
age="P90D",
)
converter.run_conversion(
nwbfile_path="sub-M001_ses-20260310.nwb",
metadata=metadata,
overwrite=True,
)
The session start time comes from the fileCreateTime field of the
.meta file, which is the local time of the acquisition computer with
no time zone attached. If you leave it that way, PyNWB assumes the time
zone of the computer running the conversion. If the date has been
zeroed out of the .meta file, as is sometimes done to de-identify
human recordings, supply a complete, time zone aware datetime
yourself. For the subject, species is the Latin binomial, sex is
one of M, F, U, or O, and age is an ISO 8601 duration.
To try the script on a small slice of the data first, ask each interface for a stub:
conversion_options = {
name: dict(stub_test=True) for name in converter.data_interface_objects
}
converter.run_conversion(
nwbfile_path="stub.nwb",
metadata=metadata,
conversion_options=conversion_options,
overwrite=True,
)
What NeuroConv Extracts and What You Supply
From the SpikeGLX files alone, NeuroConv writes:
ElectricalSeriesAPandElectricalSeriesLFin acquisition, stored as the original integers with the conversion factor to volts. When the run has more than one probe, the names gain a probe suffix (ElectricalSeriesAPImec0,ElectricalSeriesAPImec1).- An electrodes table with one row per recording site, shared by the AP and LF bands of a probe, and a device entry for each probe with its model and serial number.
- The NIDQ analog channels and the sync channel of each probe as
TimeSeries, and the NIDQ digital lines as events tables, one per line by default. - The session start time, and the sampling rate and start time of each stream.
The remainder has to come from you:
- The time zone, the subject, and the session description, as in the script above.
- The brain region of each electrode. The
locationcolumn of the electrodes table is written asunknownby default; see the NeuroConv guide on annotating extracellular electrophysiology data. - What each NIDQ channel means. The SpikeGLX page of the Conversion Gallery shows how to name analog channel groups and choose which digital lines become events.
- Trials, stimuli, and behavior, which are not in the SpikeGLX files. Each needs its own interface or a few lines of PyNWB.
Multiple Probes, Multiple Streams, and Synchronization
The converter picks up every probe in the run folder without further
configuration. If the folder holds more than one trigger file (t0,
t1, and so on), each is written as its own series with a numeric
suffix.
Each stream is written with its own sampling rate and start time.
NeuroConv takes the start time from the firstSample field of that
stream’s .meta file, so the series do not start at zero and do not
all start at the same instant.
NeuroConv does not go further than this on its own. Each probe and the NI device run on separate clocks, and NeuroConv does not use the sync square wave to correct for drift between them. The sync channels are kept in the file, but the timestamps are not adjusted. If your analysis depends on precise timing across streams over a long recording, align them before writing. There are two common routes:
- Run the SpikeGLX tools CatGT and TPrime, which extract the sync edges and map times from one stream onto another, and give the resulting times to NeuroConv.
- Align in NeuroConv. The recording interfaces have
set_aligned_timestampsandalign_by_interpolation, andSpikeGLXNIDQInterface.get_event_times_from_ttl(channel_name="XA0")returns the rising edge times of a TTL on an NIDQ channel.
imec1_ap = converter.data_interface_objects["imec1.ap"]
imec1_ap.align_by_interpolation(
unaligned_timestamps=sync_edge_times_on_imec1,
aligned_timestamps=sync_edge_times_on_imec0,
)
Do this before calling run_conversion. The same methods apply to
data recorded on another system, such as a camera. NeuroConv’s
temporal alignment documentation
describes the three supported approaches.
Adding Spike Sorting Output
KiloSortSortingInterface reads a Kilosort output folder (the one
containing params.py) directly. If you curated the sorting in Phy,
use PhySortingInterface, which reads the same folder and can leave
out clusters by curation label through exclude_cluster_groups. A
ConverterPipe puts the sorting and the raw data in the same file:
from neuroconv import ConverterPipe
from neuroconv.converters import SpikeGLXConverterPipe
from neuroconv.datainterfaces import KiloSortSortingInterface
spikeglx_converter = SpikeGLXConverterPipe(folder_path="path/to/my_run_g0")
sorting_interface = KiloSortSortingInterface(
folder_path="path/to/kilosort_output"
)
# Put the spike times on the clock of the AP stream that was sorted
ap_interface = spikeglx_converter.data_interface_objects["imec0.ap"]
sorting_interface.register_recording(ap_interface)
converter = ConverterPipe(
data_interfaces={
"SpikeGLX": spikeglx_converter,
"Kilosort": sorting_interface,
}
)
From here the script continues as the first one does: get the metadata
from converter, apply the same edits, and call run_conversion.
The register_recording call matters. Kilosort reports each spike as a
sample index counted from the start of the .ap.bin file, while the AP
series in the NWB file starts at the stream’s start time, which is
usually not zero. Without the call, spike times are counted from zero
and sit earlier than the raw trace by that amount. Any alignment of
the AP interface has to happen before the registration. The call
assumes Kilosort ran on the .ap.bin file as recorded, not on a
trimmed or concatenated copy.
The units table written this way holds the spike times and the cluster
properties Kilosort saved, such as KSLabel, but does not link each
unit to its electrodes. SortedSpikeGLXConverter adds that link from a
mapping of unit IDs to channel IDs, and handles several probes sorted
separately; see
How to Link Sorted Data to Electrodes.
Register the recording there in the same way.
Checking the File and Publishing on DANDI
Run NWB Inspector with the DANDI configuration on the result before converting the remaining sessions:
pip install nwbinspector
nwbinspector sub-M001_ses-20260310.nwb --config dandi
Our guide to sharing neurophysiology data with NWB and DANDI covers the next steps in full: viewing the file in Neurosift, creating a dandiset, uploading, and embargo.
Custom Rig Data and Other Formats
A session is usually more than its SpikeGLX files: there is often behavioral and stimulus data from a custom rig as well. CatalystNeuro builds conversion pipelines for labs. Our conversion portfolio is public, and you can book a free consultation if you would like help with yours.